User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 6032A:System Power Supply DATE: 07-Feb-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 53 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- R N P F W 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 DISP The following test equipment is required: 1.004 DISP 1. Keithley 179A DTB # 9-78 1.004 DISP 2. Transistor Devices DLP 50-150-3000A 1.004 DISP 3. Sensitive Research 50 Amp Shunt DTB # 14-3 1.005 DISP Set the UUT front panel controls as follows: 1.005 DISP LINE ON/OFF ........................................ OFF 1.005 DISP OVP ADJUST .................................... fully CW 1.006 DISP Connect the System port 1 to the UUT HP-IB connector. 1.007 DISP Remove the UUT rear panel terminal strip cover. 1.008 DISP All connections are made to the rear panel 1.008 DISP terminal strip. 1.009 DISP ÿ Connect the UUT and the 1281 as follows: 1.009 DISP ÿ[27][91]1m 6032A TO 1281[27][91]m 1.009 DISP ÿ OUTPUT (+S) ÄÄÄÄÄÄÄÄÄÄÄÄ> INPUT-HIGH 1.009 DISP ÿ OUTPUT (-S) ÄÄÄÄÄÄÄÄÄÄÄÄ> INPUT-LO 1.010 DISP Connect UUT AC line cord to Variac (220 VAC/30A). 1.011 DISP Monitor 115 AC Volts at one leg of the 220V variac with 1.011 DISP the Keithley 179A set to read AC Volts, 200 V range. 1.012 DISP Set the variac to 115 VAC. 1.013 IEEE [SRQ OFF] 1.014 DISP Set the UUT front panel controls as follows: 1.014 DISP LINE ON/OFF ......................................... ON 1.015 IEEE [D3000] 1.016 CALL HP 6032A: Sub 1 1.017 CALL HP 6032A: Sub 2 1.018 IEEE CLR; 1.019 HEAD 1.020 DISP Set the UUT front panel controls as follows: 1.020 DISP LINE ON/OFF ........................................ OFF 1.021 HEAD {} 1.022 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.022 DISP {º THIS COMPLETES THE VERIFICATION º} 1.022 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.023 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 7 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.